HD-20 Industrial Online Conductivity Transmitter Selection Instructions
| HD-20 series |
code |
Industrial online conductivity transmitter |
| Installation Method |
one |
M36×2 |
| two |
Horizontal flow pool (connecting flange DN25 or DN40) |
| three |
Right angle flow pool (connecting flange DN25 or DN40) |
| four |
Submerged installation (connecting flange DN25 or DN40) |
| five |
Water vapor circulation pool |
Main technical parameters
1. Power supply: 24VDC
2. Measurement range: 0.05-200 μ S/cm
3. Output signal: 4-20mA
4. Allowable medium pressure: standard ≤ 700Kpa (user specified if exceeding 700Kpa)
5. Temperature compensation method: automatic compensation
6. Allowable ambient temperature: 0-60 ℃
7. Medium temperature: ABS material -10-60 ℃ Stainless steel material -20-110 ℃
8. Electrode: Titanium electrode constants of 0.01 and 0.1
9. Accuracy: 1%
10. Load characteristics: 0-500 ohms
11. Installation method: flange, thread, pool through pool