DLM24 and DLM50 are new level measuring instruments manufactured with technology imported from Kay Ray Company in the United States. They are all controlled by an eight bit single-chip microcomputer, using the principle of echo method to perform non-contact continuous measurement of the height of liquid or solid materials inside the container. According to the user's needs, the instrument can display numbers in centimeters, meters, percentages, inches, feet, and other units at will. The instrument has multiple analog outputs such as 4-20mA and three sets of alarm contact outputs.
Technical Indicators:
The on-site installation of instruments is convenient, easy to adjust, and requires minimal maintenance.
The sensors of the instrument have aluminum casing, stainless steel casing, and Kynar ® There are three types of fluoroplastic shells, most of which can be equipped with internal temperature compensation and are suitable for various occasions such as general or strong corrosion.
DLM24 can be widely used to measure various crushed solid or granular materials and liquids containing a large number of particles with a diameter of not less than 2cm.
|
model |
DLM24 |
DLM50 |
|
range |
7.3m |
15m |
|
blind spot |
0.46m |
1.50m |
|
Digital display mode |
4-digit LED (display unit user selectable) |
|
Ft.、in.、m、cm、% |
Ft.、in.、m、% |
|
|
Controller |
0~45℃ |
|
|
|
|
|
Distance between sensor and controller |
90m |
200m |
|
|
|
Standard Type: LMS15KS |
LMS100R
LMS100RSS |
|
Internal temperature compensation type: LMS15KT |
LMS100RT
LMS100RTSS |
|
Low temperature type: LMS15KL (with internal heating) |
|
|
shell |
Kynar ® fluoroplastic |
cast aluminum |
|
frequency |
Approximately 42kHz |
About 14kHz |
|
Beam Angle |
Approximately 12 º cone angle |
Approximately 7 º cone angle |
|
weight |
About 0.6kg |
About 8.8kg |
Accuracy: Digital display: ± 0.5% of full range
Analog output: ± 1% of full range
Temperature additional error: ± 0.5%/10 ℃ (using temperature compensation)
Resolution: Digital display: ± 0.25% of full range
Analog output: ± 0.5% of full range
Analog output: 4-20mADC (non isolated, maximum load 750 Ω)
0~5V
0~1mA
Alarm output: three-point DPDT, contact capacity: 2A 220V AC resistive
·Method 1: Three independently adjustable relays, with action points adjustable between 2% and 99% of the range (with a switching difference of 1% of the range).
·Method 2: Two independently adjustable relays, with the operating point adjustable between 2% and 99% of the range (with a switching difference of 1% of the range). Two relays with adjustable switching differences can be set, and the operating point can be set between 2% and 99% of the range (the switching difference is the difference between the two operating points).
Power supply: 220VAC 50Hz 60W
Sensor pressure resistance: atmospheric pressure
Connection cable: The sensor and controller are connected using American RG62A/G or domestically produced SYV-75-5-1 high-frequency cables. Two core shielded cable for temperature compensation probe: It is recommended to use RVVP0.75mm2 (65 ℃) or FNP0.8mm2 (105 ℃)