Purpose and characteristics
The XF30G multifunctional standard source is an intelligent AC and DC standard voltage and current calibrator. (The communication output is displayed as true significant value). The VFD screen displays the actual output value, percentage, and output frequency, and can select the full-scale value and corresponding step size based on the range and scale of the tested meter. The amplitude adjustment can be controlled by keyboard (key) or potentiometer. Standard source, wide range, fine grading, high accuracy, good stability, easy to use, and easy to carry. Suitable for calibration, verification, and repair of five and a half digit multimeters, various pointer type AC and DC voltage and current meters below 0.1 level; It can also be used as a high stability testing power supply, in conjunction with higher-level standard meters, to calibrate current and voltage meters above level 0.05.
Main functions
The output quantity is displayed as a 6 and a half digit number, with both actual value and percentage displayed simultaneously.
Use a keyboard or potentiometer to control the increase or decrease of output. The keyboard control quantity is divided into 100%/N, 10%/N, 1%/N, and 0.1%/N (N is 4, 5, 6, 10, 15).
The external control regulator can control the increase or decrease of output quantity off the machine, and the regulator has two configurations: keyboard and potentiometer.
The output range of AC and DC voltage is 0-1000V.
The output range of AC and DC current is 0-20A.
The available AC frequencies are 50Hz, 60Hz, 100Hz, 400Hz, 500Hz, 800Hz, 1kHz (with crystal oscillator to ensure frequency stability) or mains synchronization.
AC 50Hz output can be synchronized with the mains power to reduce the frequency impact of the tested meter.
Output overload can be automatically protected and manually reset.
technical performance
Stability: AC<50ppm/5 minutes at full scale
Communication distortion:<0.5%
DC ripple factor:<0.05% (at rated output)
Output frequency accuracy: 50Hz, 60Hz, 100Hz, 400Hz, 500Hz, 800Hz, 1kHz<0.1% (or mains synchronization)
Range and accuracy of output voltage and current (23 ℃± 2 ℃, output value greater than 10% range)
|
Output project |
output range |
rated output |
Maximum Output |
accuracy |
|
AC
Voltage |
0-300mV |
20mA |
100mA |
Basic error:
DC: ± (0.01% reading)
+0.005% range)
AC: ± (0.02% reading)
+0.01% range)
ACI: ± (0.03% reading)
+0.02% range)
1kHz additional error:
± 0.03% reading |
|
0-1V-3V |
50mA |
100mA |
|
0-5V-10V-30V-100V |
200mA |
500mA |
|
0-250V-500V-1000V |
50mA |
100mA |
|
AC
current |
0-100mA |
|
0-3V-100V |
|
0-1A |
|
DC3V
AC12V |
|
0-20A |
|
1.5V |
|
Note: When the working environment temperature exceeds 23 ℃± 2 ℃, for every 10 ℃ change, the additional error is less than the basic error of the range |
Supply voltage: AC 220V ± 10%, frequency 50Hz ± 1Hz
Working environment: temperature 15 ℃~25 ℃, relative humidity 40%~70%
Working hours: continuous
Product dimensions: 180 × 480 × 480mm ³
Weight:<22kg