HomeYour location:Homeinstrumentation and metersanalytical instrumentthickness gauge → TT240 Coating thickness gauge
Column tabs 
Product name: TT240 Coating thickness gauge
specification: TT240 Coating thickness gauge
  Zoom Zoom in on the image
Category: instrumentation and meters -- thickness gauge
Column tabs 
Price: factory price
Brand: Shanghai
Place of Origin: China
Column tabs 
Available Quantity: batch
delivery cycle: Spot goods (or inquire by telephone)
tabs
  Shanghai Beiyuan Industry and Trade Co.,Ltd
+86-21-66770508
+86-13901609058

Email 91way@163.com
Wechat 13901609058(Wechat)


TT240 Coating thickness gaugeDetailed product description:Column tabs
The TT240 coating thickness gauge adopts eddy current thickness measurement method, which can measure the thickness of non-conductive coatings on non-magnetic metal substrates without damage (such as enamel, rubber, paint coatings on copper, aluminum, zinc, tin and other substrates)
·There are two measurement methods: Continuous measurement method and Single measurement method;
·It has two working modes: direct mode (Direct) and group mode (APPL);
·There are five statistical measures: Mean, Maximum, Minimum, Number of Tests, and Standard Deviation;
·Zero point calibration and two-point calibration can be performed, and basic calibration methods can be used to correct the systematic errors of the measuring head;
·Storage function: capable of storing 350 measurement values;
·It has a deletion function: it can delete individual suspicious data that appears during measurement, as well as all data in the storage area, in order to conduct new measurements;
·Limits can be set: automatic alarms can be triggered for measurement values outside the limits;
·Capable of printing: able to print measurement values and statistical values;
·Capable of communicating with a PC: capable of transmitting measured and statistical values to the PC for further data processing;
·Equipped with power undervoltage indication function;
·There is a buzzing sound during the operation process;
·Equipped with error prompt function;
·Equipped with automatic shutdown function

Probe type

N

Measuring principle

eddy current

measurement range

0-1250um/0-40um (chrome plated on copper)

Low resolution limit

1 µ m (0.1um below 10um)

Probe connection method

Split type wire connection

indication error

A little calibration (um)

±[3%H+1.5]

Two point calibration (um)

±[(1%~3%)H+1.5]

measurement conditions

Minimum curvature radius (mm)

Convex 3 concave 10

Diameter of the minimum area of the substrate (mm)

ф5

Minimum critical thickness (mm)

zero point three

temperature and humidity

0~40℃,20%RH~90%RH

Statistical Function

Mean, Max, Min
Test frequency (NO.), standard deviation (S.DEV)

working method

Direct method and group method (Appl)

measurement method

Continuous measurement method and Single measurement method

Upper and lower limit settings

Nothing

storage capacity

350 measurement values

Print/Connect Computer

Optional printer/able to connect to computer

Shutdown method

Manual and automatic

Power supply

Two AA type alkaline batteries

Overall dimensions

152×47×35mm

weight

370g