DCG-760A type electromagnetic acid-base concentration meter conductivity meterDetailed product description:

Characteristics of DCG-760A Electromagnetic Acid Base Concentration Meter Conductivity Meter:
The structural characteristics of the instrument are disk mounted, dry and wet separated structure, which is easy to install, operate and maintain
The instrument displays in a backlit dot matrix format, using a Chinese menu, while also displaying the temperature and concentration (or conductivity) of the measured solution
It has a (4-20) mA isolated output signal, and its corresponding concentration or conductivity output upper and lower limits can be set arbitrarily
Equipped with conductivity upper limit alarm and concentration upper and lower limit alarm functions
Equipped with power-off data protection function, even if there is an abnormal power outage, the data stored inside the instrument is still valid. After restarting, it can enter the measurement state without calibration
Measure the conductivity of a solution using electromagnetic induction, without polarization phenomenon, with accurate measurement and easy maintenance
The instrument is equipped with a single-chip microcomputer inside, which can accurately compensate for the temperature of the solution
It can measure the concentration of HC1 solution or NaOH solution, as well as the conductivity of general solutions
The protection level of the electronic unit chassis is IP65, which can prevent water spray and dust intrusion
Equipped with RS-232 or RS-485 output interface, capable of digital output
Supporting transmitter: DCGF-100A electromagnetic transmitter
Technical specifications of DCG-760A electromagnetic acid-base concentration meter conductivity meter:
1. Measurement range:
a) Concentration: (0-10.0)% HCI
(0~10.0)%NaOH
b) Conductivity: (0-50) mS/cm
(0~200)mS/cm
(0~500)mS/cm
(0~1000)mS/cm
c) Temperature: (-10.0~120) ℃
2. Basic error of the instrument:
a) Concentration: ± 0.3% HCI ± one word
± 0.3% NaOH ± one word
b) Conductivity: 2.0% (FS) ± one word
c) Temperature: ± 0.5 ℃± one word (0.0 ℃≤ T ≤ 60.0 ℃)
± 1.0 ℃ ± one word (in other ranges)
3. Temperature of the tested solution: (5.0-45.0) ℃
4. Temperature coefficient compensation range: (0-5.0)%/℃ (adjustable)
5. Stability of electronic units:
a) Concentration: ± 0.1% HCl/24h
±0.1%NaOH/24h
b) Conductivity: ± 0.3% FS/24h
6. Water sample inlet and outlet connection: Φ 8 (inner diameter) × 12 reinforced polyvinyl chloride (PVC) hose
7. Maximum operating temperature of measuring probe: 55.0 ℃
8. Maximum operating pressure of measuring probe: 0.5 × 106Pa
9. Dimensions of electronic unit, mm: 144 × 144 × 150
10. Electronic unit opening size, mm: 138 × 138
11. Weight: Approximately 2kg
12. The cable length of the measuring probe is generally 10 meters. If an additional length is required, it must be specifically stated when ordering