SPB200 precision digital pressure gauge is an improved low-power intelligent instrument. It adopts advanced low-power devices and sophisticated software technology, with a built-in lithium battery that can work continuously for 5-8 years without the need to connect to an external power source; It adopts a large screen wide viewing angle segment code LCD display window, with very eye-catching and intuitive readings. The appearance is available in radial, axial, and disc mounting options.
Application
Used for verifying pressure (differential pressure) transmitters, precision pressure gauges, ordinary pressure gauges, blood pressure monitors, and other pressure instruments
Accuracy reaches 0.05 level, meeting the requirements of high-precision pressure acquisition
Meet the pressure collection requirements of corrosive, impact, vibration and other places
Replace pointer precision pressure gauge and can be used as a standard pressure gauge
Meet the measurement and testing standard instrument requirements of various industries
Characteristics of SPB200 Precision Digital Pressure Gauge
Measurement accuracy up to 0.05 level
Design of magnetic pen induction button
Intuitively display pressure percentage bar chart
The collection speed can be set
Anti electromagnetic and radio frequency interference technology
Power management technology, battery life can reach 5-8 years, lifetime battery replacement free
Zero point self stabilization technology
Temperature compensation technology
Technical data
measurement range
Gauge pressure: -100kPa~0~1kPa~260MPa
Differential pressure: 0~500Pa~3.5MPa
Absolute pressure: 0~10kPa~60MPa
Accuracy level: 0.05 level, 0.1 level, 0.2 level, 0.5 level
Display method: 5-digit dynamic display, percentage bar chart
Overload pressure: 1.5-3 times the range, depending on the measurement range
Built in power supply: 3.6V high-energy lithium battery
Medium temperature: -40~85 ℃
Work environment:
Temperature: -30 ℃ to 60 ℃; Humidity:<90%
Temperature: 0 ℃~50 ℃ (0.05 level operating temperature); Humidity:<90%
Calibration reference environment: 20 ℃± 2 ℃
Explosion proof level: ExiaIIBT4
Process connection: M20 × 1.5 (external thread) or according to user requirements
Optional accessory: High end aluminum alloy carrying case
Dimensions: (Unit: mm)