SYD-2122A Automatic Trace Moisture Tester (Automatic Coulombic Method)Detailed product description:

The SYD-2122A Automatic Trace Moisture Tester (Automatic Coulombic Method) is designed and manufactured in accordance with the requirements of the Chinese standard GB 7600 "Determination of Moisture Content in Operating Transformers (Coulombic Method)" and the Chinese industry standard SH/T0246 "Determination of Moisture Content in Light Petroleum Products (Electric Method)". It is suitable for determining the content of trace moisture in liquid petroleum products according to the methods specified in the above standards.
The SYD-2122A Automatic Trace Moisture Tester (Automatic Coulombic Method) has the characteristics of fast measurement speed, high accuracy, accurate and reliable data, and simple operation. It can be applied in departments such as petroleum, chemical, power, medicine, environmental protection, geology, etc.
1、 Main technical features
1. This instrument adopts microcontroller control technology, which enables seamless sampling, testing, calculation, analysis, and result output, with a high degree of automation.
2. This instrument has the characteristics of fast measurement speed, high accuracy, accurate and reliable data, and simple operation.
3. The instrument is a desktop structure with a simple design, easy to use, and complete supporting facilities. Purchasing the instrument allows for testing of trace amounts of moisture.
2、 Main technical parameters and indicators
1. Titration method: Electric titration;
2. Electrolytic current: 0-300mA;
3. Measurement range: 10 μ g to 30mg;
4. Sensitivity threshold: 1 μ g H2O;
5. Accuracy: 10 μ g to 1mg H2O ± 5 μ g, with no less than 0.5% for amounts greater than 1mg H2O;
6. Power supply: AC220V ± 10%, frequency 50Hz ± 5%;
7. Power consumption: 25VA.