HomeYour location:Homeinstrumentation and metersSpecialized instruments and metersPetroleum instruments → SYD-267 Petroleum product open flash point and ignition point tester
Column tabs 
Product name: SYD-267 Petroleum product open flash point and ignition point tester
specification: SYD-267 Petroleum product open flash point and ignition point tester
  Zoom Zoom in on the image
Category: instrumentation and meters -- Petroleum instruments
Column tabs 
Price: factory price
Brand: Shanghai
Place of Origin: China
Column tabs 
Available Quantity: batch
delivery cycle: Spot goods (or inquire by telephone)
tabs
  Shanghai Beiyuan Industry and Trade Co.,Ltd
+86-21-66770508
+86-13901609058

Email 91way@163.com
Wechat 13901609058(Wechat)


SYD-267 Petroleum product open flash point and ignition point testerDetailed product description:Column tabs
1、 Purpose:
The SYD-267 petroleum product open flash point and ignition point tester shall comply with the requirements specified in the national standard GB/T267 "Determination of Flash Point and Ignition Point of Petroleum Products (Open Cup Method)" and the national petrochemical industry standard SH/T0318 "Technical Conditions for Open Flash Point Testers". Determine the flash point and ignition point of lubricating oil and dark petroleum products.

2、 Technical parameters
1. Working power supply: AC 220V ± 10%, 50Hz.
2. Outer crucible ⑴ Material: 0.3 high-quality carbon structural steel, surface coated with black; ⑵ Inner diameter of upper opening: Φ 100 ± 5 millimeters; ⑶ Height: 50 ± 5 millimeters; Bottom inner diameter: Φ 56 ± 2 millimeters.
3. Inner crucible ⑴ Material: 0.3 high-quality carbon structural steel, surface coated with black;
⑵ Inner diameter of upper opening: Φ 64 ± 1 mm;
⑶ Height: 47 ± 1 millimeters;
Bottom inner diameter: 38 ± 1 mm;
⑸ Scale line: There is a scale line at a distance of 12 millimeters and 18 millimeters from the upper edge.
4. Gas duct: The nozzle diameter is between 0.8 and 1 millimeter, and the smooth surface of the inner hole can adjust the flame by 3-4 millimeters.
5. Heater: Electric furnace heating, adjustable heating power of 1000W.
6. Temperature gauge: 0-360 ℃, with a division of 1 ℃, in accordance with the technical specifications of GB/T514 for liquid thermometers used in petroleum product testing.
8. Environmental temperature: ≤ 35 ℃.
9. Relative humidity: ≤ 85%.
10. Power consumption of the whole machine: not exceeding 1100W.

3、 Complete set of instruments
SYD-267 Open Flash Point Tester
Inner crucible
External crucible
Igniter
Temperature gauge 0-360 ℃
Temperature gauge clamp (column)
Power cord (250V 6A)
Fuse 5A (Ø 5 × 20)
SYD-267 Petroleum product open flash point and ignition point tester
Product Category
All product information on this website is translated by Baidu machine translation, which may not be completely accurate and is only for reference