Product name: |
SYD-267 Petroleum product open flash point and ignition point tester
|
|
specification: |
|

Zoom in on the image |
|
Category: |
instrumentation and meters
-- Petroleum instruments |
|
Price: |
factory price |
|
Brand: |
Shanghai |
|
Place of Origin: |
China |
|
Available Quantity: |
batch |
|
delivery cycle: |
Spot goods (or inquire by telephone) |
|
| |
Shanghai Beiyuan Industry and Trade Co.,Ltd
+86-21-66770508
+86-13901609058 |
91way@163.com
13901609058(Wechat) |
|
|
|
SYD-267 Petroleum product open flash point and ignition point testerDetailed product description:

1、 Purpose:
The SYD-267 petroleum product open flash point and ignition point tester shall comply with the requirements specified in the national standard GB/T267 "Determination of Flash Point and Ignition Point of Petroleum Products (Open Cup Method)" and the national petrochemical industry standard SH/T0318 "Technical Conditions for Open Flash Point Testers". Determine the flash point and ignition point of lubricating oil and dark petroleum products.
2、 Technical parameters
1. Working power supply: AC 220V ± 10%, 50Hz.
2. Outer crucible ⑴ Material: 0.3 high-quality carbon structural steel, surface coated with black; ⑵ Inner diameter of upper opening: Φ 100 ± 5 millimeters; ⑶ Height: 50 ± 5 millimeters; Bottom inner diameter: Φ 56 ± 2 millimeters.
3. Inner crucible ⑴ Material: 0.3 high-quality carbon structural steel, surface coated with black;
⑵ Inner diameter of upper opening: Φ 64 ± 1 mm;
⑶ Height: 47 ± 1 millimeters;
Bottom inner diameter: 38 ± 1 mm;
⑸ Scale line: There is a scale line at a distance of 12 millimeters and 18 millimeters from the upper edge.
4. Gas duct: The nozzle diameter is between 0.8 and 1 millimeter, and the smooth surface of the inner hole can adjust the flame by 3-4 millimeters.
5. Heater: Electric furnace heating, adjustable heating power of 1000W.
6. Temperature gauge: 0-360 ℃, with a division of 1 ℃, in accordance with the technical specifications of GB/T514 for liquid thermometers used in petroleum product testing.
8. Environmental temperature: ≤ 35 ℃.
9. Relative humidity: ≤ 85%.
10. Power consumption of the whole machine: not exceeding 1100W.
3、 Complete set of instruments
SYD-267 Open Flash Point Tester
Inner crucible
External crucible
Igniter
Temperature gauge 0-360 ℃
Temperature gauge clamp (column)
Power cord (250V 6A)
Fuse 5A (Ø 5 × 20)

Related Products: