HomeYour location:Homeinstrumentation and metersSpecialized instruments and metersPetroleum instruments → SYD-8018D gasoline oxidation stability tester (induction period method)
Column tabs 
Product name: SYD-8018D gasoline oxidation stability tester (induction period method)
specification: SYD-8018D gasoline oxidation stability tester (induction period method)
  Zoom Zoom in on the image
Category: instrumentation and meters -- Petroleum instruments
Column tabs 
Price: factory price
Brand: Shanghai
Place of Origin: China
Column tabs 
Available Quantity: batch
delivery cycle: Spot goods (or inquire by telephone)
tabs
  Shanghai Beiyuan Industry and Trade Co.,Ltd
+86-21-66770508
+86-13901609058

Email 91way@163.com
Wechat 13901609058(Wechat)


SYD-8018D gasoline oxidation stability tester (induction period method)Detailed product description:Column tabs
1、 Main technical features
1. The SYD-8018D gasoline oxidation stability tester (induction period method) is a desktop structure that utilizes various advantages of a PC to provide the whole machine with a series of advantages such as simple operation, precise control, and high degree of automation.
2. The test host is connected to the computer using the most advanced USB interface, making it easy to use.
3. Change the traditional water bath to a metal bath, which is pollution-free, does not require water replenishment, and is more convenient to operate and use.
4. A well-designed insulation system not only saves energy but also avoids the risk of operator burns.
5. The integrated design of the testing oxygen bomb assembly shortens the oil and gas pressure testing pipeline, making pressure detection more accurate, and ensuring good sealing of the oxygen bomb during the testing process, making the test results more reliable.
6. During the testing process, various parameters of the system were displayed in real-time on the computer interface, and the pressure curves inside the two oxygen bombs were also drawn in real time, making the test results clear at a glance.

2、 Main technical parameters and indicators
1. Working power supply: AC220V ± 5%, 50Hz.
2. Heating tube power: 1600W, actual heating power is automatically controlled by the computer.
3. Oxygen bomb pressure transmitter measurement range: 0-1600kPa, accuracy: ± 2 ‰.
4. Temperature control point for metal bath: 100.0 ℃± 1 ℃.
5. Temperature gauge: Equipped with a glass mercury thermometer, the system can be calibrated as needed at any time.
6. Environmental temperature: ≤ 30 ℃.
7. Relative humidity: ≤ 85%.
SYD-8018D gasoline oxidation stability tester (induction period method)