WN10 endotoxin detectorDetailed product description:

WN10 endotoxin detector
The WN10 endotoxin detector, as required by the 2002 edition of the Chinese Pharmacopoeia for bacterial endotoxin testing, states that the reaction tube should be placed vertically in a suitable thermostat at 37 ± 1 ℃ and kept warm for 60 ± 2 minutes. According to this requirement, our company has developed a special experimental equipment for bacterial endotoxin test by gel method - WN10 endotoxins detection thermostat. Compared with the traditional constant temperature water bath, the WN10 type test tube thermostat has the following advantages and characteristics:
1. The WN10 endotoxin detector adopts the most advanced heating technology and pulse temperature control technology, with a temperature control accuracy of 37 ± 0.15 ℃ and a temperature difference between test tube holes of ≤ 0.3 ℃. The temperature control accuracy of the traditional constant temperature water bath is only ± 1 ℃, and the temperature distribution is uneven. The water temperature near the heating tube is high, and the water temperature away from the heating tube is low, which can easily cause differences in the reaction of horseshoe crab reagents.
2. The endotoxin detector WN10 is equipped with a timing device, and users can adjust the timing time arbitrarily within 0-99H59 minutes; And there is an independent buzzing sound. 50 test tube wells of the instrument can be used simultaneously.
3. The instrument is equipped with a dust cover, and the test tube is not sealed during use, making experimental operations more convenient and efficient. If using a traditional constant temperature water bath, the test tubes must be sealed, otherwise water vapor or small water droplets will enter the test tubes and cause contamination. Sealing each test tube one by one is not only troublesome and time-consuming, but also affects the accuracy of the experiment.
4. The WN10 endotoxin detector has been calibrated to a constant temperature before leaving the factory, and users do not need to adjust the instrument temperature. Just turn on the instrument preheating 15-20 minutes before each use, and the instrument temperature will be displayed on the screen.
5. The endotoxin detector WN10 has 50 test tube wells per instrument. The instrument can be customized with 2ml ampoules as reaction tubes and 10 × 75mm standard reaction tubes according to user needs. The instrument has a beautiful appearance, compact size, and is convenient to use.
Main Features
1. The WN10 endotoxin detector displays real-time temperature and decreasing time
2. Automatic fault detection and alarm function
3. Buzzer alarm prompt function
4. Temperature deviation calibration function
5. Convenient module replacement for easy cleaning and disinfection
6. Built in over temperature protection device
7. It has interrupt and pause function
8. LCD display, membrane switch
Product parameters
1. Temperature control range of endotoxin detector WN10: room temperature+5-100 ℃
2. Heating time: ≤ 20 minutes (from 20 ℃ to 100 ℃)
3. Temperature control accuracy: ≤± 0.5 ℃
4. Display accuracy: 0.1 ℃
5. Module temperature uniformity: ≤± 0.5 ℃
6. Maximum power: 200W
7. Maximum temperature: 100 ℃
8. Time setting: up to 99 hours and 59 minutes