Visible spectrophotometer
Model: V-5100
The instrument adopts a 128 * 64 dot matrix LCD display, which can directly display standard curves and test data, with clear display, accurate readings, stable and reliable performance
Can continuously test and store 200 sets of data, with 5 sets displayed on each screen
In C mode (standard curve method), it is possible to directly establish multi-point standard curves and use the established standard curves for unknown sample concentration testing
Can store 200 standard curves, and can be called up freely according to their numbers
In F-mode (coefficient method), the coefficients of the curve equation can be input and directly measured for the sample
The measured and stored data has power-off hold function
Automatic wavelength calibration, automatic setting, and deviation self repair
Socket type tungsten lamp design, no optical debugging required for lamp replacement
Equipped with standard USB data output interface and parallel print output interface, convenient for online operation and printing data
The quantitative software from Yuanxi Company can directly complete photometric analysis, quantitative testing, and data processing
Instrument specifications
| model |
V-5100 |
| wavelength range |
325-1000nm |
| Spectral bandwidth |
4nm |
| Wavelength accuracy |
±2nm |
| wavelength repeatability |
≤0.5nm |
| Wavelength setting method |
automatic |
| wavelength resolution |
0.1nm |
|
Photometric accuracy |
±0.5%T |
| Photometric repeatability |
≤0.2%T |
| stray light |
≤0.05%T |
| stability |
± 0.001A/30min (after preheating at 500nm) |
| Luminous range |
0-200%T、-0.3-3A、0-9999C |
| data output |
USB port |
| print output |
Parallel port |
| display system |
128 * 64 bit large screen LCD |
| light source |
Imported long-life tungsten lamp |
| Overall dimensions |
420*280*180mm |
| Power supply |
AC 85V~250V |
| weight |
10kg |