| Product name: | MP521 Laboratory pH/ Conductivity Meter | ||||||
| specification: | ![]() |
||||||
| Category: | instrumentation and meters -- Conductivity Meter | ||||||
| Price: | factory price | ||||||
| Brand: | Shanghai | ||||||
| Place of Origin: | China | ||||||
| Available Quantity: | batch | ||||||
| delivery cycle: | Spot goods (or inquire by telephone) | ||||||
|
|||||||
|
pH |
measurement range
resolution
accuracy
input current
input impedance
stability
Temperature compensation range |
(-1.99 ~ 19.99)pH
0.1/0.01 pH
Electricity meter: ± 0.01 pH, matching: ± 0.02 pH
≤1×10-12 A
≥3×1012 Ω
±0.01 pH/3h
(0~100) ℃ (automatic or manual) |
|
mV |
Measurement range (mV/ORP/EH)
resolution
accuracy |
-1999mV ~ 0 ~ 1999mV
1mV
±0.1% FS |
|
conductivity |
measurement range
resolution
accuracy
Temperature compensation range
electrode constant
Reference temperature |
Conductivity: (0~200) mS/cm, divided into five ranges: (0.00~20.00) μ S/cm (20.0~200.0) μ S/cm (200~2000) μ S/cm
(2.00~20.00)mS/cm (20.0~200.0)mS/cm
Electrical resistivity: (0~100) M Ω· cm
TDS: (0 ~ 100)g/L
Salinity: (0~100) ppt
0.01/0.1/1μS/cm 0.01/0.1 mS/cm
Electricity meter: ± 1.0% FS, matching: ± 1.50% FS
(0~50) ℃ (automatic)
0.1 / /1 / 10 / cm-1
25℃ |
|
temperature |
measurement range
resolution
accuracy |
-10℃~ 110℃
0.1℃
5~60 ℃ range: ± 0.5 ℃ Other ranges: ± 1 ℃ |
|
Other technical parameters |
Data storage
Storage Content
Power supply
communication interface
Dimensions and Weight
Quality and Safety Certification |
900 sets
Measurement value number, measurement value, temperature value, ATC or MTC status, measurement date, measurement time
DC9V/300mA
RS232
160 × 190 × 70mm/880g
ISO9001:2000, CE and CMC |
|
working conditions |
Ambient temperature
Ambient Humidity
IP rating |
5 ~ 35 ℃
≤85%
IP54 dustproof and splash proof |