Main features of DDS-307A conductivity meter
Adopting a large screen, blue backlight, dual row digital display LCD
Equipped with manual and automatic temperature compensation functions
Novel graphic design, appearance design patent number: ZL 200430036029.2
Electrical connectors are lower than the plane, which helps with corrosion prevention and safety
Selecting a titanium alloy electrode with a conductivity cell constant of 0.01cm-1 and a sealed measurement slot can be suitable for high-purity water measurement
Equipped with our factory's DJS-1C conductivity electrode (platinum black) and T-818-B-6 temperature electrode
Technical Indicators:
Instrument level: 1.0 level
1. Measurement range:
Conductivity: (0-1 × 105) μ S/cm
TDS:(0~1000) mg/L
Temperature: (0-60) ℃
2. Basic error:
Conductivity: ± 0.5% (FS) ± 1 word
Temperature: ± 0.5 ℃
3. Stability: ± 0.33% (FS) ± 1 word/3h
4. Temperature compensation range: (0-35) ℃; Reference temperature 25 ℃
5. Power supply: AC (220 ± 22) V; (50 ± l) Hz
6. Dimensions (mm): 300 × 200 × 72
7. Weight: 1.5kg
8. The conductivity cell constant corresponds to the optimal conductivity measurement range
|
Electrode constant cm-1 |
zero point zero one |
zero point one |
one |
ten |
|
Measurement range μ S/cm |
0~2 |
0.2~20 |
2~1×10four |
1×10four~1×10five |