The main features of DDS-304 conductivity meter are:
Dual use AC/DC, meter type reading
Equipped with manual calibration, constant manual compensation, and temperature manual compensation functions
Selecting a titanium alloy electrode with a conductivity cell constant of 0.01cm-1 and a sealed measurement slot can be suitable for high-purity water measurement
Equipped with DJS-1C conductivity electrode (platinum black)
Main technical specifications of DDS-304 conductivity meter:
Instrument level: 2.0 level
1. Measurement range:
Conductivity: (0-1 × 105) μ S/cm, divided into seven measurement ranges; When the electrode constant is 0.01 μ S/cm-1
Range: 0-0.1, 0-0.3, 0-1, 0-3, 0-10, 0-30, 0-100, 0-100
2. Basic error: Conductivity: ± 1.5% (FS)
3. Stability: ± 1% (FS)/3h
4. Manual temperature compensation range: (15-35) ℃; Reference temperature 25 ℃
5. Power supply: Can be equipped with CD-965K (9V, 300mA) power supply; GNB0.28K-4.8V nickel cadmium battery (1 cell) can also be used
6. Dimensions (mm): 210 × 140 × 80
7. Weight: Approximately 1kg
8. The conductivity cell constant corresponds to the optimal conductivity measurement range
|
Electrode constant cm-1 |
zero point zero one |
zero point one |
one |
ten |
|
Measurement range μ S/cm |
0~2 |
0.2~20 |
2~1×10four |
1×10four~1×10five |