| technical indicators |
| DC voltage |
| Range: |
0 - 600.0 V |
| Accuracy: |
1% ± 5 words | |
| AC voltage |
| Fluke 333/334 |
|
| Range: |
0 - 600.0 V |
| Accuracy: |
1% ± 5 words (50/60 Hz) |
| Communication response: |
average |
| Fluke 335 |
|
|
|
| Range: |
0 - 600.0 V |
| Accuracy: |
1% ± 5 words (50/60 Hz) |
| Communication response: |
True effective value |
|
|
| Fluke 336/337 |
|
| Range: |
0 - 600.0 V |
| Accuracy: |
1% ± 5 words (20-100 Hz), 6% ± 5 words (100-400 Hz) |
| Communication response: |
True effective value | |
| direct current |
| Fluke 336 |
|
| Range: |
0 - 600.0 A |
| Accuracy: |
2% ± 3 words |
|
|
| Fluke 337 |
|
| Range: |
0 - 999.9 A |
| Accuracy: |
2% ± 3 words | |
| alternating current |
| Fluke 333/334 |
|
| Range: |
0 - 400.0 A |
| Accuracy: |
2% ± 5 words (50/60 Hz) |
| Communication response: |
average |
|
|
| Fluke 335 |
|
| Range: |
0 - 600.0 A |
| Accuracy: |
2% ± 5 words (50/60 Hz) |
| Peak coefficient (50/60 Hz): |
2.4 @ 500A, 2.0 @ 600A, When CF>2, increase by 2% |
| Communication response: |
True effective value |
|
|
| Fluke 336 |
|
| Range: |
0 - 600.0 A |
| Accuracy: |
2% ± 5 words (10-100 Hz), 6% ± 5 words (100-400 Hz) |
| Peak coefficient (50/60 Hz): |
3 @ 500A, 2.5 @ 600A, When CF>2, increase by 2% |
| Communication response: |
True effective value |
|
|
| Fluke 337 |
|
| Range: |
0 - 999.9 A |
| Accuracy: |
2% ± 5 words (10-100 Hz), 6% ± 5 words (100-400 Hz) |
| Peak coefficient (50/60 Hz): |
3 @ 500A, 2.5 @ 600A, 1.42@1000 A, When CF>2, increase by 2% |
| Communication response: |
True effective value | |
| resistance |
| Fluke 333 |
|
| Range: |
0 - 600.0 Ω |
| Accuracy: |
1.5% ± 5 words |
|
|
| Fluke 334/335/336/337 |
|
| Range: |
0 - 600.0 Ω, 6001 - 6000 Ω |
| Accuracy: |
1.5% ± 5 words | |
| frequency |
| Fluke 337 |
|
| Range: |
5 - 400Hz |
| Accuracy: |
0.5% ± 5 words |
| Trigger Level: |
10-100 Hz >= 5A, 5-10 Hz, 100-400 Hz >=10A | |
| Inrush current |
| Fluke 334/335/336/337 |
|
| Integral time: |
100 ms | |
| continuity |
|
| MIN/MAX |
|
Only applicable to Fluke 337 model | |