MS2307,MS2306 Intelligent Grounding Resistance MeterDetailed product description:Column tabs
Detailed information

1) Six measurement methods, including bipolar AC, bipolar DC, three pole, four pole, current clamp selection method, and dual clamp head without auxiliary pole method, are used to measure the grounding resistance, meeting the needs of various measurement environments.  
2) The measurement voltage is optional at 48V/20V, and the measurement frequency is available in four options: 94Hz, 105Hz, 111Hz, and 128Hz. It also has an automatic frequency selection function, which automatically selects the frequency with the least interference for measurement.
3) The 4-pole method can be used to directly measure soil resistivity;
4) Capable of automatically measuring interference voltage and leakage current;
5) It has a measurement data storage function, which can store 100 sets of measurement data and upload them to a PC for display analysis and output test reports.
6) Dual power supply, optional dry battery and rechargeable battery power supply technical specifications:

Intelligent Grounding Resistance Tester MS2307 Hong Kong Huayi Parameters:

measurement method
function
MS2308
MS2307
MS2306
Measurement of Three pole and Four level Grounding Resistance
Test voltage: AC20V/48V
Test frequency: 94Hz/105Hz/111Hz/128Hz/AFC
94Hz/128Hz
Range of grounding resistance
0.02Ω~300KΩ
0.1Ω~30KΩ
0.1Ω~20KΩ
Selection method test (current clamp)
Test voltage: AC20V/48V
nothing
Test frequency: 94Hz/105Hz/111Hz/128Hz/AFC
Range of grounding resistance
0.02Ω~20KΩ
0.1Ω~10KΩ
Non auxiliary pole method (voltage clamp+current clamp)
Test voltage: AC48V
nothing
Test frequency: 94Hz/105Hz/111Hz/128Hz/AFC
Range of grounding resistance
0.02Ω~150Ω
0.1Ω~150Ω
Measurement of bipolar AC resistance
Test voltage: AC20V
Test frequency: 94Hz/105Hz/111Hz/128Hz/AFC
94Hz/128Hz
Range of grounding resistance
0.02Ω~300KΩ
0.1Ω~30KΩ
0.1Ω~20KΩ
Two pole/four pole DC resistance measurement
Test voltage: DC20V
nothing
Range of grounding resistance
0.02Ω~3KΩ
0.1Ω~3KΩ
Interference voltage/interference current/interference frequency
Interference voltage 1-50V DC/AC
Interference frequency 16Hz~400Hz
Interference current 20mA~2A
Soil resistivity
Test voltage: AC20V/48V
Test frequency: 94Hz/105Hz/111Hz/128Hz/AFC
94Hz/128Hz
Range: 0.02 Ω· m~1000K&Omega,
RK (Test Wire Compensation)
Test voltage: AC20V/48V
Test frequency: 94Hz/105Hz/111Hz/128Hz/AFC
94Hz/128Hz
Range:
0.02~30Ω
0.1~30Ω
0.1~30Ω
test current
≦250mA
≦100mA
≦100mA
data storage
USB communication

MS2307,MS2306 Intelligent Grounding Resistance Meter