HomeYour location:Homeinstrumentation and metersanalytical instrumentBiochemical Analysis → EN-610 Hydrogen Analyzer (Split)
Column tabs 
Product name: EN-610 Hydrogen Analyzer (Split)
specification: EN-610 Hydrogen Analyzer (Split)
  Zoom Zoom in on the image
Category: instrumentation and meters -- Biochemical Analysis
Column tabs 
Price: factory price
Brand: Shanghai
Place of Origin: China
Column tabs 
Available Quantity: batch
delivery cycle: Spot goods (or inquire by telephone)
tabs
  Shanghai Beiyuan Industry and Trade Co.,Ltd
+86-21-66770508
+86-13901609058

Email 91way@163.com
Wechat 13901609058(Wechat)


EN-610 Hydrogen Analyzer (Split)Detailed product description:Column tabs
The EN-610 thermal conductivity hydrogen analyzer operates based on the physical property that different gases have different thermal conductivities, and the thermal conductivity of mixed gases varies with the content of the measured components.
The EN-610 thermal conductivity hydrogen analyzer adopts stable and reliable thermal conductivity gas sensors and microprocessor technology. The instrument also has functions such as digital display, upper and lower limit alarm, standard signal output, and relay contact alarm output. Gas online analysis suitable for industrial production and other analytical occasions.

Key Features:
The instrument adopts a thermal conductivity gas sensor
Full Chinese operation menu
The instrument adopts a large screen LCD display screen
Real time clock display
Can be connected to a printer to achieve timed automatic printing
Equipped with timed automatic storage function and the ability to view stored data at any time
Equipped with RS232 communication port, it can be connected to a computer to achieve real-time data transmission, data storage, curve trend chart, printing report and other functions
Can set upper and lower limit alarm points, with alarm contact output
Online calibration with available standard gas
The output current can be selected as 0-10mA or 4-20mA

Main technical parameters:
Measurement range: 0-2%, 0-10%, 35-75%, 50-80%, 80-100% (volume)
Accuracy: Level 2.5
Response time: T90 < 40 seconds
Output signal: 4-20mADC (allows external load<800 Ω) 0-10mADC (allows external load<1500 Ω)
Power supply: 220V ± 10% AC frequency 50Hz ± 5%
Environmental temperature: 0 ℃~45 ℃
Environmental humidity: 5-90% RH. 50 ℃
Power consumption: 60W
Sample gas flow rate: 200 milliliters per minute
Gas consumption: approximately 30l/h
Controller dimensions: 144 × 144 × 250mm
Controller opening size: 138 × 138mm
Analyzer dimensions: 500 × 360 × 250mm
Weight: Approximately 4.5Kg

EN-610 Hydrogen Analyzer (Split)
Product Category
All product information on this website is translated by Baidu machine translation, which may not be completely accurate and is only for reference