ZO-802 Zirconia Oxygen AnalyzerDetailed product description:

Characteristics of ZO-802 Zirconia Oxygen Analyzer:
High quality zirconia detection components are selected, which have the characteristics of long lifespan and fast response.
Adopting a large screen dot matrix LCD display, it can simultaneously display parameters such as oxygen content, temperature, date, and time.
Adopting a fully Chinese human-machine dialogue menu, the operation is intuitive and convenient.
Adopting a convection diffusion structure reduces the influence of flow rate on the measurement of trace oxygen.
The instrument has eight ranges from trace to constant, with automatic range conversion.
The upper and lower limit alarm points can be set arbitrarily within the full range.
The instrument has an automatic storage function for measurement data, and the storage interval time is set by the user.
Equipped with a paperless recorder function, it automatically records the curve of oxygen concentration changes over time.
Has 0~10 or 4~20mA current signal output.
The instrument has a standard RS232C communication port, which can be connected to a serial printer or achieve bidirectional communication with a computer.
The instrument adopts a small flow meter, making it more compact.
Technical specifications of ZO-802 zirconia oxygen analyzer:
Measurement range: 0.1ppm~50% O2
Automatic range shifting: 0.1~10ppm, 0~100ppm, 0~1000ppm
0~1%, 0~5%, 0~10%, 0~25%, 0~50%, a total of 8 gears (automatic shifting).
Basic error: ± 3% FS (>100ppm); ±5%FS(≤100ppm)
Repetitive error: ≤ 1.5%
Response time: T90 < 30 seconds
Zero drift: ± 3% FS/7d (>100ppm); ±5%FS/7d(≤100ppm)
Range drift: ± 3% FS/7d (>100ppm); ±5%FS/7d(≤100ppm)
Output signal: 4-20mADC (allows external load<800 Ω)
0-10mADC (allows external load<1600 Ω)
Alarm relay contact capacity: 220VAC/1A
Power consumption: ≤ 50W
Power supply: 220V ± 10% AC; 50Hz ± 5%
Environmental temperature: 0~40 ℃
Relative humidity: not exceeding 90% (excluding condensation)
Atmospheric pressure: local atmospheric pressure
Sample gas temperature: 0~40 ℃
Sample gas pressure: < 0.2Mpa
Sample gas flow rate: 300ml/min
Instrument dimensions: 240 × 150 × 280mm (width × height × depth)
Installation hole size: 228 × 138mm
Weight: 8kg

Related Products: