EN-630 Type CO2 AnalyzerDetailed product description:

The EN-630 thermal conductivity gas analyzer is a physical analyzer that can be used to analyze the content of a certain component (referred to as the test component) in a gas mixture. It applies the physical property of changes in the content of the component to be tested in a gas mixture. The change in thermal conductivity leads to a change in the resistance value of the hot wire, indirectly determining the content of the component to be tested. In the process of heat conduction, different solids have different thermal conductivity coefficients, resulting in different rates of heat conduction. Those with higher thermal conductivity coefficients transfer heat faster, while those with lower thermal conductivity transfer heat slower. Gases, like solids, can conduct heat, and the speed of heat transfer is also related to the thermal conductivity of various gases. So the EN-630CO2 analyzer works based on this principle. The instrument adopts a stable and reliable thermal conductivity gas sensor and microprocessor technology, with functions such as large screen LCD display, upper and lower limit alarm, standard signal output, and relay contact alarm output. Used for measuring carbon dioxide concentration, it is also suitable for online gas analysis in industrial production and other analytical applications.
Key Features:
The instrument adopts a thermal conductivity gas sensor
Full Chinese operation menu
The instrument adopts a large screen LCD display screen
Real time clock display
Can be connected to a printer to achieve timed automatic printing
Equipped with timed automatic storage function and the ability to view stored data at any time
Equipped with a paperless recorder function, it automatically records the curve of carbon dioxide concentration over time
Equipped with RS232 communication port, it can be connected to a computer to achieve real-time data transmission, data storage, curve trend chart, printing report and other functions
Can set upper and lower limit alarm points, with alarm contact output
Online calibration with available standard gas
The output current can be selected as 0-10mA or 4-20mA
Main technical parameters:
Measurement range: 0~10% 0~20% 0~30% CO2/Ar (volume)
Basic error: ± 1.5%
Response time: T90 < 40 seconds
Output signal: 4-20mA DC (allows external load<800 Ω)
0-10mA DC allows external load<1600 Ω)
Power supply: 220V ± 10% AC frequency 50Hz ± 5%
Environmental temperature: 0-40 ℃
Environmental humidity: 5-90% RH
Power consumption: 60W
Sample gas flow rate: 200 milliliters per minute
Inlet temperature: 0-40 ℃
Instrument dimensions: 144 × 144 × 250mm
Instrument opening size: 138 × 138mm
Weight: Approximately 4.5Kg

Related Products: