HomeYour location:Homeinstrumentation and metersSpecialized instruments and metersCalibration instrument → CA100 Multi functional process calibrator
Column tabs 
Product name: CA100 Multi functional process calibrator
specification: CA100 Multi functional process calibrator
  Zoom Zoom in on the image
Category: instrumentation and meters -- Calibration instrument
Column tabs 
Price: factory price
Brand: Shanghai
Place of Origin: China
Column tabs 
Available Quantity: batch
delivery cycle: Spot goods (or inquire by telephone)
tabs
  Shanghai Beiyuan Industry and Trade Co.,Ltd
+86-21-66770508
+86-13901609058

Email 91way@163.com
Wechat 13901609058(Wechat)


CA100 Multi functional process calibratorDetailed product description:Column tabs
Technical specifications of CA100 multifunctional process calibrator

Functional specifications occur
DC voltage occurs, DC current occurs, and resistance occurs

function
scope
Accuracy guarantee range
resolution
display range
Precision
±(%of setting+%of range)
Temperature coefficient * 1
±(%of setting+%of range)/℃
Remark
%of setting
%of range
%of setting/℃
%of range/℃
DC voltage occurrence
100mV
0~100.000mV
1μV
-10~110.000mV
zero point zero two
zero point zero one
zero point zero zero three
zero point zero zero two
Output resistance: approximately 6.5 Ω
1V
0~1.00000V
10μV
-0.1~1.10000V
zero point zero two
zero point zero zero five
zero point zero zero two
zero point zero zero one
Maximum output: 10mA
Output resistance: approximately 30m Ω
10V
0~10.0000V
100μV
-1~11.0000V
zero point zero two
zero point zero zero five
zero point zero zero two
zero point zero zero one
Maximum output: 10mA
Output resistance: approximately 30m Ω
DC voltage occurrence
20mA
0~20.000mA
1μA
0~22.000mA
zero point zero two five
zero point zero one five
zero point zero zero three
zero point zero zero three
Maximum output: 24V
-20mA*2
0~-20.000mA
1μA
0~-22.000mA
zero point zero two five
zero point zero three
zero point zero zero three
zero point zero zero three
Maximum output: 28V
Resistance occurrence * 3
500Ω
0~500.00Ω
10mΩ
0~550.00Ω
0.02*4
0.02*4
zero point zero zero two
zero point zero one
*45
5kΩ
0~5.0000kΩ
100mΩ
0~5.5000Ω
0.05*4
0.03*4
zero point zero zero two
zero point zero one
*46
50kΩ
0~50.000kΩ
0~55.000Ω
0.1*4
0.1*4
zero point zero zero two
zero point zero three
*47

The temperature conditions for the temperature coefficient are: 5 ℃ to 18 ℃ and 28 ℃ to 40 ℃.
DC-mA suction: The function of drawing a specified current value in the positive direction from an external voltage source.
Resistance occurrence: Similar to the equivalent resistance occurrence method, its specifications are only valid within the range of measured current and output voltage in the remarks column.
Excluding the accessory wire resistance part.
Measure current of 1mA to 5mA and output voltage below 2V.
Measure current of 0.1mA to 1mA and output voltage below 2V.
Measure current ranging from 0.01mA to 0.1mA and output voltage below 2V.

Frequency occurrence, pulse occurrence

function
scope
(Accuracy guarantee range 0)
display resolution
Display range of occurrence
Precision
maximum output
Remark
Frequency occurrence
1~100Hz
100.0Hz
1~110Hz
±1digits
10mA
Waveform: Rectangular wave with approximately 50% duty
Output level: 0-10V
Output level accuracy: ± 10%
100~1000Hz
1000 Hz
90~1100Hz
1~10kHz
10.0k Hz
0.9~11.0kHz
10~50kHz
50k Hz
9~50kHz
Pulse occurrence * 8
1~100Hz
60,000cycles
1~60000cycles
±1digits
10mA
100~1000Hz
1~10kHz
10~50kHz

Thermocouple potential occurrence

scope
Temperature range of occurrence ℃
Resolution ℃
Accuracy ℃
Temperature coefficient * 9 (℃/℃)
K
-200.0~-100.0
zero point one
zero point six
zero point zero five
-100.0~400.0
zero point one
zero point five
zero point zero five
400.0~1200.0
zero point one
zero point seven
zero point zero five
1200.0~1372.0
zero point one
zero point nine
zero point zero five
E
-250.0~-200.0
zero point one
one point two
zero point one
-200.0~-100.0
zero point one
zero point six
zero point zero five
-100.0~600.0
zero point one
zero point five
zero point zero five
600.0~-1000.0
zero point one
zero point six
zero point zero five
J
-210.0~-100.0
zero point one
zero point six
zero point zero five
-100.0~-800.0
zero point one
zero point five
zero point zero five
800.0~1200.0
zero point one
zero point seven
zero point zero five
T
-250.0~-200.0
zero point one
one point five
zero point two
-200.0~400.0
zero point one
zero point five
zero point zero five
N
-200.0~-100.0
zero point one
one
zero point one
-100.0~900.0
zero point one
zero point seven
zero point zero five
900.0~1300.0
zero point one
zero point eight
zero point zero five
B
400~600
one
two
zero point two
600~800
one
one point five
zero point two
800~1820
one
one point one
zero point two
R
-40~100
one
one point five
zero point two
100~1767
one
one point two
zero point two
S
-40~100
one
one point five
zero point two
100~1767
one
one point five
zero point two

Temperature conditions for temperature coefficient: 5 ℃ to 18 ℃ and 28 ℃ to 40 ℃
This specification complies with IEC 584-1-1989 and IEC 584-1-1995
The internal resistance of the thermocouple output is approximately 6.5 Ω
The accuracy does not include RJC accuracy, and standard contact compensation is performed using RJC sensors sold separately.
When using standard contact temperature for output compensation, RJC accuracy should be added.
Output compensation is performed approximately every 10 seconds
RJC sensor specifications:
Measurement range: -10~50C
Accuracy: 18-28 ℃: ± 0.5 ℃ (combined with the main body)
-10-18 ℃, 28-50 ℃: ± 1 ℃ (combined with the main body)
Wire length: approximately 1.5m
Temperature unit: ℃, F

Thermal resistance output occurs:

scope
Temperature range of occurrence ℃
Resolution ℃
Accuracy * 10 ℃
Temperature coefficient * 11 ℃
PT100
-200.0~0
zero point one
zero point three
zero point zero four
0~400.0
zero point one
zero point five
zero point zero four
400.0~850.0
zero point one
zero point eight
zero point zero four

The accuracy is calculated based on the measured current of 1-5mA, excluding the auxiliary wire resistance part.
The temperature conditions for the temperature coefficient are: 5 ℃ to 18 ℃ and 28 ℃ to 40 ℃.
This specification complies with IEC 751-1983 and IEC 751-1995

General specifications for signal generation function
The output of the thermistor adopts the equivalent resistance generation method.
Response time of the department:
DC voltage occurrence, within the range of 1V or 10V, 10ms (time from output change to entering accuracy)
Resistance occurrence, 500 Ω range (including thermal resistance output): 10ms (the time from the specified power supply to the output entering the accuracy range)
Functions occurring outside the above range: 300ms
Maximum load capacity: 0.01 μ F (occurrence of DC voltage, resistance, thermal resistance EMF, RTD frequency, and pulse)
Minimum load capacity: 100 μ F (DC current generated)
Voltage limiter of the generator: operates when the load is above 28.5V (manual reset)
Current limiter of the generator: Load action above 12mA (limited to DC voltage, frequency, pulse occurrence, manual reset)
Temperature unit: ℃, F

General specifications for signal generation function
Output=set value x n/m
When m=integer (1 15); N=integer (1 m)

24V DC power supply function
Output voltage: 24 ± 1V
Maximum output current: 22mA
Output protection, output stops at maximum 30mA

Measurement function specifications

function
scope
Accuracy guarantee range
resolution
display range
Precision
±(%of reading+%of range)
Temperature coefficient * 1
±(%of reading+%of range)/℃
Remark
%of reading
%of range
(%of setting)/℃
(%of range)/℃
DC voltage occurrence
500mV
-500.00~500.00mV
10μV
-599.99~599.99mV
zero point zero two
zero point zero one
zero point zero zero two
zero point zero zero one
Input resistance: 1G Ω or above
5V
-5.0000~5.0000V
100μV
-5.9999~5.9999V
zero point zero two
zero point zero one
zero point zero zero two
zero point zero zero one
Input resistance: approximately 1m Ω
35V
-35.000~35.000V
1mV
-41.999~41.999V
zero point zero two
zero point zero one five
zero point zero zero two
zero point zero zero one
DC voltage occurrence
20mA
-20.000~20.00mA
1μA
-23.999~23.999mA
zero point zero two five
zero point zero two
zero point zero zero two
zero point zero zero one
Input resistance: below 20 Ω
100mA
-100.00~100.00mA
10μA
-119.99~119.99mA
zero point zero four
zero point zero three
zero point zero zero two
zero point zero zero one
Resistance occurrence * 3
500Ω
0~500.00Ω
10mΩ
0~599.99Ω
0.055*13
0.015*13
zero point zero zero five
zero point zero two
Measurement current: approximately 1mA
5kΩ
0~5.0000kΩ
100mΩ
0~5.9999Ω
0.055*13
0.015*13
zero point zero zero five
zero point zero two
Measurement current: approximately 100 μ A
50kΩ
0~50.000kΩ
0~59.999Ω
0.55*13
0.02*13
zero point zero zero five
zero point zero two
Measurement current: approximately 10 μ A

The temperature conditions for the temperature coefficient are: 5 ℃ to 18 ℃ and 28 ℃ to 40 ℃.
Resistance without accessory wires
CA100 Multi functional process calibrator
Product Category
All product information on this website is translated by Baidu machine translation, which may not be completely accurate and is only for reference