function |
scope |
Accuracy guarantee range |
resolution |
display range |
Precision
±(%of setting+%of range) |
Temperature coefficient * 1
±(%of setting+%of range)/℃ |
Remark | ||
|
%of setting |
%of range |
%of setting/℃ |
%of range/℃ | ||||||
|
DC voltage occurrence |
100mV |
0~100.000mV |
1μV |
-10~110.000mV |
zero point zero two |
zero point zero one |
zero point zero zero three |
zero point zero zero two |
Output resistance: approximately 6.5 Ω |
|
1V |
0~1.00000V |
10μV |
-0.1~1.10000V |
zero point zero two |
zero point zero zero five |
zero point zero zero two |
zero point zero zero one |
Maximum output: 10mA
Output resistance: approximately 30m Ω | |
|
10V |
0~10.0000V |
100μV |
-1~11.0000V |
zero point zero two |
zero point zero zero five |
zero point zero zero two |
zero point zero zero one |
Maximum output: 10mA
Output resistance: approximately 30m Ω | |
|
DC voltage occurrence |
20mA |
0~20.000mA |
1μA |
0~22.000mA |
zero point zero two five |
zero point zero one five |
zero point zero zero three |
zero point zero zero three |
Maximum output: 24V |
|
-20mA*2 |
0~-20.000mA |
1μA |
0~-22.000mA |
zero point zero two five |
zero point zero three |
zero point zero zero three |
zero point zero zero three |
Maximum output: 28V | |
|
Resistance occurrence * 3 |
500Ω |
0~500.00Ω |
10mΩ |
0~550.00Ω |
0.02*4 |
0.02*4 |
zero point zero zero two |
zero point zero one |
*45 |
|
5kΩ |
0~5.0000kΩ |
100mΩ |
0~5.5000Ω |
0.05*4 |
0.03*4 |
zero point zero zero two |
zero point zero one |
*46 | |
|
50kΩ |
0~50.000kΩ |
1Ω |
0~55.000Ω |
0.1*4 |
0.1*4 |
zero point zero zero two |
zero point zero three |
*47 | |
|
function |
scope
(Accuracy guarantee range 0) |
display resolution |
Display range of occurrence |
Precision |
maximum output |
Remark |
|
Frequency occurrence |
1~100Hz |
100.0Hz |
1~110Hz |
±1digits |
10mA |
Waveform: Rectangular wave with approximately 50% duty
Output level: 0-10V
Output level accuracy: ± 10% |
|
100~1000Hz |
1000 Hz |
90~1100Hz | ||||
|
1~10kHz |
10.0k Hz |
0.9~11.0kHz | ||||
|
10~50kHz |
50k Hz |
9~50kHz | ||||
|
Pulse occurrence * 8 |
1~100Hz |
60,000cycles |
1~60000cycles |
±1digits |
10mA | |
|
100~1000Hz | ||||||
|
1~10kHz | ||||||
|
10~50kHz |
|
scope |
Temperature range of occurrence ℃ |
Resolution ℃ |
Accuracy ℃ |
Temperature coefficient * 9 (℃/℃) |
|
K |
-200.0~-100.0 |
zero point one |
zero point six |
zero point zero five |
|
-100.0~400.0 |
zero point one |
zero point five |
zero point zero five | |
|
400.0~1200.0 |
zero point one |
zero point seven |
zero point zero five | |
|
1200.0~1372.0 |
zero point one |
zero point nine |
zero point zero five | |
|
E |
-250.0~-200.0 |
zero point one |
one point two |
zero point one |
|
-200.0~-100.0 |
zero point one |
zero point six |
zero point zero five | |
|
-100.0~600.0 |
zero point one |
zero point five |
zero point zero five | |
|
600.0~-1000.0 |
zero point one |
zero point six |
zero point zero five | |
|
J |
-210.0~-100.0 |
zero point one |
zero point six |
zero point zero five |
|
-100.0~-800.0 |
zero point one |
zero point five |
zero point zero five | |
|
800.0~1200.0 |
zero point one |
zero point seven |
zero point zero five | |
|
T |
-250.0~-200.0 |
zero point one |
one point five |
zero point two |
|
-200.0~400.0 |
zero point one |
zero point five |
zero point zero five | |
|
N |
-200.0~-100.0 |
zero point one |
one |
zero point one |
|
-100.0~900.0 |
zero point one |
zero point seven |
zero point zero five | |
|
900.0~1300.0 |
zero point one |
zero point eight |
zero point zero five | |
|
B |
400~600 |
one |
two |
zero point two |
|
600~800 |
one |
one point five |
zero point two | |
|
800~1820 |
one |
one point one |
zero point two | |
|
R |
-40~100 |
one |
one point five |
zero point two |
|
100~1767 |
one |
one point two |
zero point two | |
|
S |
-40~100 |
one |
one point five |
zero point two |
|
100~1767 |
one |
one point five |
zero point two |
|
scope |
Temperature range of occurrence ℃ |
Resolution ℃ |
Accuracy * 10 ℃ |
Temperature coefficient * 11 ℃ |
|
PT100 |
-200.0~0 |
zero point one |
zero point three |
zero point zero four |
|
0~400.0 |
zero point one |
zero point five |
zero point zero four | |
|
400.0~850.0 |
zero point one |
zero point eight |
zero point zero four |
|
function |
scope |
Accuracy guarantee range |
resolution |
display range |
Precision
±(%of reading+%of range) |
Temperature coefficient * 1
±(%of reading+%of range)/℃ |
Remark | ||
|
%of reading |
%of range |
(%of setting)/℃ |
(%of range)/℃ | ||||||
|
DC voltage occurrence |
500mV |
-500.00~500.00mV |
10μV |
-599.99~599.99mV |
zero point zero two |
zero point zero one |
zero point zero zero two |
zero point zero zero one |
Input resistance: 1G Ω or above |
|
5V |
-5.0000~5.0000V |
100μV |
-5.9999~5.9999V |
zero point zero two |
zero point zero one |
zero point zero zero two |
zero point zero zero one |
Input resistance: approximately 1m Ω | |
|
35V |
-35.000~35.000V |
1mV |
-41.999~41.999V |
zero point zero two |
zero point zero one five |
zero point zero zero two |
zero point zero zero one | ||
|
DC voltage occurrence |
20mA |
-20.000~20.00mA |
1μA |
-23.999~23.999mA |
zero point zero two five |
zero point zero two |
zero point zero zero two |
zero point zero zero one |
Input resistance: below 20 Ω |
|
100mA |
-100.00~100.00mA |
10μA |
-119.99~119.99mA |
zero point zero four |
zero point zero three |
zero point zero zero two |
zero point zero zero one | ||
|
Resistance occurrence * 3 |
500Ω |
0~500.00Ω |
10mΩ |
0~599.99Ω |
0.055*13 |
0.015*13 |
zero point zero zero five |
zero point zero two |
Measurement current: approximately 1mA |
|
5kΩ |
0~5.0000kΩ |
100mΩ |
0~5.9999Ω |
0.055*13 |
0.015*13 |
zero point zero zero five |
zero point zero two |
Measurement current: approximately 100 μ A | |
|
50kΩ |
0~50.000kΩ |
1Ω |
0~59.999Ω |
0.55*13 |
0.02*13 |
zero point zero zero five |
zero point zero two |
Measurement current: approximately 10 μ A | |