| Product name: | YD9880 series programmable safety comprehensive tester | ||||||
| specification: | ![]() |
||||||
| Category: | instrumentation and meters -- Safety Compliance Tester | ||||||
| Price: | factory price | ||||||
| Brand: | Shanghai | ||||||
| Place of Origin: | China | ||||||
| Available Quantity: | batch | ||||||
| delivery cycle: | Spot goods (or inquire by telephone) | ||||||
|
|||||||
| Withstand Voltage Test | ||
|
AC |
Output voltage range |
0.5kV-5kV |
|
Voltage resolution |
10V | |
|
Voltage accuracy |
3% | |
|
voltage frequency |
50Hz/60Hz optional | |
|
Maximum output current |
40mA | |
|
Current setting |
0.01mA-40mA | |
|
Current Resolution |
1μA | |
|
Current accuracy |
3% | |
|
DC |
Output voltage range |
0.5kV-6kV |
|
Voltage resolution |
10V | |
|
Voltage accuracy |
3% | |
|
Maximum output current |
10mA | |
|
Current setting |
0.01mA-10mA | |
|
Current Resolution |
1μA | |
|
Current accuracy |
3% | |
| Insulation resistance test | ||
|
IR |
Voltage measurement range |
100V-1000V |
|
Voltage resolution |
1V | |
|
Voltage accuracy |
± (3% y reading+3V) | |
|
Resistance measurement range |
1MΩ-9999MΩ | |
|
Accuracy of resistance measurement |
10% | |
|
Resistance upper limit setting |
1MΩ-9999MΩ | |
|
Lower limit setting of resistance |
1MΩ-9999MΩ | |
| Grounding resistance test | ||
|
GR |
Output voltage range |
3V-8V AC |
|
voltage frequency |
50Hz/60Hz optional | |
|
Output current range |
3A-30A AC | |
|
Current Resolution |
0.01A | |
|
Current accuracy |
3% | |
|
Resistance testing range |
1mΩ-600mΩ | |
|
Resistance display resolution |
1mΩ | |
| Accuracy of resistance measurement |
3% | |
|
Resistance setting range |
1mΩ-600mΩ(3-10A) 1mΩ-150mΩ(330A) | |
| Leakage current test (only YD9881, YD9882, YD9880A are passive leakage) | ||
|
LC |
Voltage display range |
100V-250V |
| Voltage display resolution |
1V | |
| Voltage display accuracy |
3% | |
| Current measurement range |
0.01mA-20mA | |
| Current Resolution |
0.01mA(2-20)mA 0.01mA(0.0012)mA | |
| Accuracy of current display |
3% | |
| Current upper limit setting |
0.01mA-20mA | |
|
Isolation transformer capacity |
2000VA | |
| Leakage testing method |
hot state | |
| Power test (YD9882 only) | ||
|
PW |
Voltage display range |
0-250V |
| Voltage display resolution |
1V | |
| Current display range |
0-20A | |
| Current display resolution |
0.01A | |
| Power display range |
0-3000W | |
| Power display resolution |
1W | |
| Measurement accuracy |
3% | |
| Isolation transformer capacity |
2000VA | |
| Start Test (YD9882 only) | ||
|
Run |
Voltage display range |
0V-250V |
| Voltage display resolution |
1V | |
| Current display range |
0-20A | |
| Current display resolution |
0.01A | |
| Measurement accuracy |
3% | |
| Other technical indicators | ||
|
Monitor |
240 × 128 LCD display | |
|
Measurement time setting |
1.0-999.9s | |
|
Memory group |
1-50 groups (each group can store 4 parameters) | |
|
communication interface |
RS232C, PLC interface | |
|
Operating Temperature and Humidity |
0℃ -40℃,≤ 85%RH | |
|
Power requirements |
198V-242V AC,47.5Hz-52.5Hz | |
|
Overall dimensions |
450×250×500(mm) | |
|
weight |
About 30kg | |