Main technical parameters
1. Test force: MHV-5Z: 2.942, 4.903, 9.807, 19.61, 24.52, 29.42, 49.03N
(0.3、 0.5、 1、 2、 2.5、 3、 5 kgf)
MHV-10Z :2.942、 4.903、 9.807、 19.61、 24.52、 29.42、 49.03、 98.07N
(0.3、 0.5、1、 2、2.5、 3、 5 、10 kgf)
2. Repeatability of hardness tester readings:
|
Range of standard block hardness |
Repeatability of hardness tester readings |
|
HV5~HV100 |
HV0.2~<HV5 |
<HV0.2 |
|
E |
≤6 |
≤12 |
≤12 |
|
>225HV |
≤4 |
≤8 |
≤10 |
3. Test force application method: automatic loading, holding, unloading
4. Objective lens and indenter switching mode: automatic switching
5. Equipped with built-in printer
6. Micrometer eyepiece: 10 ×
7. Objective lens: 10 ×, 40 ×
8. Total magnification: 100x (observation), 400x (measurement)
9. Minimum detection unit: Total magnification: 100x 0.25 μ m 400x 0.0625 μ m
10. Test force holding time: 0-60 seconds (every 5 seconds as a unit)
11. Maximum height of specimen: 170 mm
12. Distance from the center of the pressure head to the machine wall: 130 mm
13. Host weight: approximately 50 kg
14. Power supply: (110~220) AV, (50~60) Hz
15. Dimensions (length x width x height): 530 x 225 x 630 mm