HomeYour location:Homeinstrumentation and metersanalytical instrumentturbidimeter → TSC-10(E) online turbidity/SS monitoring instrument
Column tabs 
Product name: TSC-10(E) online turbidity/SS monitoring instrument
specification: TSC-10(E) online turbidity/SS monitoring instrument
  Zoom Zoom in on the image
Category: instrumentation and meters -- turbidimeter
Column tabs 
Price: factory price
Brand: Shanghai
Place of Origin: China
Column tabs 
Available Quantity: batch
delivery cycle: Spot goods (or inquire by telephone)
tabs
  Shanghai Beiyuan Industry and Trade Co.,Ltd
+86-21-66770508
+86-13901609058

Email 91way@163.com
Wechat 13901609058(Wechat)


TSC-10(E) online turbidity/SS monitoring instrumentDetailed product description:Column tabs
Overview
The TSC-10 (E)/TCS-1000 (E) online turbidity/SS monitoring instrument complies with the national standard JJG 880-2006 and the international standard ISO7027-84, and is designed based on the principle of light scattering. Due to the use of infrared light sources and 90 degree scattered light measurement, interference from the color of the water being measured and sunlight can be eliminated. So, it has high measurement accuracy and good stability.

Purpose:
It is particularly suitable for online turbidity measurement in various containers and liquid flow fields such as water channels, wells, water tanks, and water tanks. It can also be widely used for accurate measurement of turbidity in various water samples in sewage treatment plants, water plants, power plants, wastewater treatment plants, pharmaceutical plants, and hospitals.

Key Features:
Never wear out lens, using artificially synthesized crystals with a hardness of 9 on the Mohs scale
Turbidity measurement technology using infrared monochromatic light scattering, with the addition of a chromaticity compensation system, effectively avoids interference from experimental colors
High limit, low limit alarm (LED light, delay) and switch control (240V/1A)
Signal output: 4~20mA output with upper and lower limits set arbitrarily
Equipped with an automatic cleaning device, eliminating the hassle of manual cleaning
One or more calibration points to ensure measurement accuracy
Switching between multiple units including NTU, FTU, and mg/l
Equipped with environmental temperature and humidity self checking function and error prompt function

Technical parameters of TSC-10 (E) online turbidity/SS monitoring instrument:

1. Measurement range: 0.00~500 NTU 0.00~1000mg/L
2. Accuracy (relative error): ≤ 10%
3. Light source: Infrared light-emitting diode (LED)
4. Detector: Integrated photodetector
5. Operating temperature: (3-40) ℃
6. Humidity usage: not exceeding 90% (at 50 ℃)
7. Sensor cable length:<9m
8. Sensor head size: diameter 48mm, length 133m
9. Power supply: AC 110~220V
10. Host size (length x width x height): 162mm x 240mm x 75mm
11. Host weight: approximately 0.5kg