HomeYour location:Homeinstrumentation and metersRecording instrumentPaperless recorder → VX6000R Color screen paperless recorder
Column tabs 
Product name: VX6000R Color screen paperless recorder
specification: VX6000R Color screen paperless recorder
  Zoom Zoom in on the image
Category: instrumentation and meters -- Paperless recorder
Column tabs 
Price: factory price
Brand: Shanghai
Place of Origin: China
Column tabs 
Available Quantity: batch
delivery cycle: Spot goods (or inquire by telephone)
tabs
  Shanghai Beiyuan Industry and Trade Co.,Ltd
+86-21-66770508
+86-13901609058

Email 91way@163.com
Wechat 13901609058(Wechat)


VX6000R Color screen paperless recorderDetailed product description:Column tabs
VX6000R Color Screen Paperless Recorder

VX6000R 16 channel universal input color screen paperless recorder can input signals such as standard current, standard voltage, frequency, millivolts, thermocouples, and thermal resistors. It has sensor isolation distribution output, relay alarm output, transmission output, flow accumulation, temperature and pressure compensation, historical data transfer, printing, and remote communication functions.

Features
system
Using the latest large-scale integrated circuits and carefully designed components from internationally renowned factories;
Adopting a high-speed and high-performance 32-bit ARM microprocessor, the screen response time is less than 0.2 seconds, and it can simultaneously detect, record, display, and alarm 12 signals
5.6-inch 320 × 234 dot matrix TFT high brightness color graphic LCD display, LED backlight, clear image, wide viewing angle. The screen brightness can be freely adjusted.
Fully isolated universal input, capable of inputting multiple signals without the need for jumpers, can be configured through software;
New switching power supply, working normally within the range of 85VAC~265V A C;
Integrated hardware real-time clock, the clock can run accurately even in the event of power failure;
Provide transmitter+24VDC isolated power distribution.
The all aluminum sealed shell and internal shielding plate ensure the normal operation of the instrument in harsh environments.
Large capacity FLASH flash memory chips store historical data and never lose data in case of power failure.
12 relay alarm outputs.

Signal
Can input various standard signals, Fr、20mA、5V、S、B、K、T、E、J、、R、N、Pt100、Cu50、Cu53、BA1、BA2、350 Ω
20mV、100mV、10mA、WRE、F1、F2
Signal full range accuracy ± 0.2%
Complete isolation between channels
Can provide isolated power distribution
Provide standard 4-20mA output

software
Chinese menu configuration, display project tag number, and engineering unit. Software password lock ensures user configuration security;
The quantity display has a wide range of numerical values and can display 5-digit values: -9999 to 19999;
Alarm display, indicating the lower and lower limits, lower and upper limits, and upper and upper limits of each channel alarm, and can record the last 15 alarm messages;
Support traffic accumulation function;
4 sets of temperature and pressure compensation, supporting compensation signal input and optional constants, providing multiple compensation models, such as commonly used compensation models for superheated steam, saturated steam pressure compensation, etc;
Support vacuum Scientific notation display;
The curved paper feeding method can be set horizontally or vertically;
Every 4 sets of curves can be freely combined and the color of the curves can be customized;
Built in GB2312 second level Chinese character library (6500 Chinese characters);
The all-new T6 input method is easy to operate. Support Chinese Pinyin input, with optional input of numbers, English, special symbols, subscripts, etc., using international standards
Standard coding, solving the input problem of special units and Chinese character numbering;
Custom combination input of Chinese character tags and units.
Communication
Standard serial communication interfaces: RS-485 and RS-232C;
Supports standard Modbus RTU communication protocol, providing multiple data types such as percentage, engineering quantity, cumulative quantity, etc. In addition to supporting ScanX data management software, it also supports popular professional configuration software such as iFlX, Kingview, MCGS, PCAUTo, etc., without the need for dedicated drivers;
Use USB 2.0 interface to transfer and backup historical records, supporting 256MB, 512MB, 1G USB drives;
Supports FAT16/FAT32 file system types, Windows can recognize history files.
External micro printer, capable of manually printing data and curves, and automatically timed printing of data to meet users' on-site printing needs.

Technical specifications of VX6000R color screen paperless recorder
monitor:
5.6-inch 320 × 234 dot matrix TFT graphic LCD display screen, LED backlight, wide viewing angle
Basic error:
±0.2%F·S
Fully isolated universal input: 1-2 channel signal input, fully isolated between channels, isolation voltage>400V
Output specifications:
Distribution: 4 sets of 24VCD output, output current 60mA per set
12 configurable relay contact outputs, with a contact capacity of 3A and 250VAC (resistive load), default to normally open contacts.
Up to 4 analog outputs (transmission output or PID output), 4-20mADC, load resistance<600 Ω
Compensation calculation:
Steam: According to the steam density table, query the steam density and compensate for the mass flow rate and pressure of the steam
Range 0.1-22 Mpa
Temperature range 0-600 ℃
Water: According to the water density table, check the water density and compensate for the mass flow rate of water within the temperature range of 0-150 ℃
Cumulative range:
0~9999999999
Communication and printing:
Communication interface: RS232C or RS485
Baud rates: 1200, 2400, 1800, 9600, 19200, 38400, 57600
Printing interface: RS232C directly connected to serial micro printer
Record time:
Record interval: 1, 2, 5, 10, 15, 30, 60, 120, 240 seconds optional.
Record length: 72 hours/pen (record interval of 1 second) to 720 days/pen (record interval of 4 minutes)
Data backup and transfer:
256MB~1G USB drive optional;
Cold end compensation error of thermocouple:
±1℃
Power outage protection:
Built in FLASH memory, protects parameters and historical data, and can be permanently saved after power failure. Integrated hardware clock that can operate accurately even after power failure.
Power supply: Switching power supply 85VAC~265VAC, 50Hz ± 5%, power consumption<30W
Environmental temperature: 0-50 ℃
Environmental humidity: 0-85% RH