| Product name: | RK1156 Six in One Comprehensive Tester | ||||||
| specification: | ![]() |
||||||
| Category: | instrumentation and meters -- Safety Compliance Tester | ||||||
| Price: | factory price | ||||||
| Brand: | Shanghai | ||||||
| Place of Origin: | China | ||||||
| Available Quantity: | batch | ||||||
| delivery cycle: | Spot goods (or inquire by telephone) | ||||||
|
|||||||
|
endure |
output voltage |
Communication between 200-5000V |
|
|
Output voltage error |
± (set value * 2%+3V) |
||
|
Measurement range of breakdown current |
0.02~20mA |
||
|
Measurement accuracy of breakdown current |
± (measured value * 3%+3V) |
||
|
Time setting range |
1 to 300 seconds |
||
|
cease |
output voltage |
DC 200-1000V |
|
|
Output voltage error |
± (set value * 2%+3V) |
||
|
Measurement range of insulation resistance |
1.0~49.9MΩ~1000MΩ |
||
|
Measurement accuracy of insulation resistance |
Below 50M Ω ± (measured value * 5%+0.5M Ω) |
||
|
Above 50M Ω ± (measured value * 10%+10M Ω) |
|||
|
pick up |
Time setting range |
1 to 300 seconds |
|
|
Output current |
Exchange 10-25A |
||
|
Output current error |
± (2% * reading value) |
||
|
Measurement range of grounding resistance |
11-25A gear, 1-200m Ω |
||
|
Measurement accuracy of grounding resistance |
± (2% reading value+3m Ω) |
||
|
Time setting range |
1 to 300 seconds |
||
|
leak |
output voltage |
1.03 or 1.11 times the input voltage (output by isolation transformer) |
|
|
Output voltage error |
± (2% reading value+3V) |
||
|
Measurement range of insulation resistance |
0.05~20.00mA |
||
|
Measurement accuracy of insulation resistance |
± (measured value * 2%+3mA) |
||
|
Time setting range |
2-300 seconds |
||
|
merit |
output voltage |
1.0 times the input voltage (output by isolation transformer) |
|
|
Voltage measurement accuracy |
± (reading value * 2%+3V) |
||
|
Current measurement range |
0.1A~20.00A |
||
|
Accuracy of current measurement |
± (measured value * 2%+3mA) |
||
|
Power measurement range |
0-6000W (or 2000W, 4000W) |
||
|
Power measurement accuracy |
± (measured value * 2%+3W) |
||
|
Time setting range |
1 to 300 seconds |
||
|
Start |
output voltage |
0.80 or 0.85 or 0.9 times the input voltage (output by isolation transformer) |
|
|
Voltage measurement accuracy |
± (measured value * 2%+3V) |
||
|
Current measurement range |
0.1A~20.00A |
||
|
Accuracy of current measurement |
± (measured value * 2%+0.03A) |
||
|
Time setting range |
1 to 300 seconds |
||
|
whole |
working power supply |
AC 200V±10%,50Hz±5% |
|
|
Whole machine no-load power consumption |
About 30W |
||
|
Usage environment |
AC 200V ± 10 ℃, relative humidity: ≤ 60% RH |
||
|
Overall dimensions |
46*58*116cm |
||
|
weight |
About 70kg |
RK1156 Six in One Comprehensive Tester