Product Description
The XF30Q-2 clamp meter calibrator is an intelligent AC and DC standard voltage and current calibrator. (The communication output is displayed as true significant value). Three windows simultaneously display the output actual value, percentage, and checked table full value. It is possible to set the full-scale value of the inspected table and select the corresponding step size based on the scale of the inspected table. The amplitude adjustment can be controlled by keyboard (key) or potentiometer. It has the characteristics of digital display, error direct reading, wide range, fine grading, high accuracy, good stability, and easy use. Suitable for calibration and verification of various AC/DC clamp meters below level 1, four and a half digit multimeters, and various 0.5 level pointer AC/DC voltage and current meters; It can also be used as a high stability testing power supply, in conjunction with high-level standard meters, to calibrate 0.1 level current and voltage meters. Main function: 5 1/2 digit LED digital display output, displaying both actual value and percentage simultaneously; Use keyboard to control the increase or decrease of output volume. The keyboard control quantity is divided into 100%/N, 10%/N, 1%/N, and 0.1%/N (N is 4, 5, 6, 10, 15); The external control regulator can control the increase or decrease of output quantity off the machine, and there are two configurations on the regulator: keyboard and potentiometer, which can be converted freely; The output range of AC and DC voltage is 0-750V; the output range of AC and DC current is 0-10A; there are four AC frequencies to choose from: 50Hz, 60Hz, 400Hz; Output overload can be automatically protected and manually reset; Clamp meter measurement: equipped with our standard coil, it can measure current from 0 to 2000A with an error of 0.3%.
technical indicators
Stability: AC<0.02% of full scale/3 minutes; DC<0.01% of full scale/3 minutes
Communication distortion:<0.5%
DC ripple factor:<0.1% (at rated output)
Output frequency accuracy: 50Hz, 60Hz, 400Hz<1%
Range and accuracy of output voltage, current, and median resistance (23 ℃± 5 ℃, output value greater than 10% range)
|
Output project |
output range |
rated output |
maximum output |
accuracy |
|
AC
Voltage |
0-200mV |
20mA |
100mA |
Basic error:
DC: ± (0.03% reading)
+0.02% range)
AC: ± (0.05% reading)
+0.03% range)
Voltage additional error: ± 0.02mV
ACI: Accuracy not assessed for ≤ 1mA |
|
0-1V-2V |
50mA |
100mA |
|
0-5V-10V-20V |
100mA |
200mA |
|
0-50V-100V-200V |
50mA |
100mA |
|
0-500V-750V |
DC20mA
AC40mA |
DC40mA
AC60mA |
|
AC
current |
0-10A |
|
DC4V
AC10V |
|
Median resistance |
0Ω-10MΩ |
0.25W |
|
±0.2%+20mΩ |
|
Note: When the working environment temperature exceeds 23 ℃± 2 ℃, for every 10 ℃ change, the additional error is less than this range
Basic error |
Power consumption: AC power supply voltage 220V ± 10%, frequency 50Hz ± 1Hz;
Working environment: The temperature of the working environment is between 5 ℃ and 35 ℃, and the relative humidity is less than 80%
Working hours: continuous
Dimensions: 180 × 480 × 480mm
Weight: Approximately 22kg