PC-WZ Series Three Channel Universal Input Paperless RecorderDetailed product description:Column tabs
The PC-WZ series three channel universal input paperless recorder can input signals such as standard current, standard voltage, frequency, millivolts, thermocouples, and thermal resistors. It has functions such as isolated distribution output, relay alarm output, transmission output, flow accumulation, temperature and pressure compensation, historical data transfer, printing, and remote communication. Through wired and wireless communication networking, a remote monitoring system can be formed with professional configuration software such as MCGS, Citect, iFIX, and Kingview.
The instrument adopts a 32-bit ARM microprocessor with powerful display functions, implementing trend curve display, historical curve recall, over limit alarm, alarm status display, and providing intelligent historical trend automatic playback function. It is extremely convenient to search for alarm points and also provides positioning playback function, directly querying data of a certain year, month, day, hour, minute, and second, which is convenient and fast.

Technical parameters of PC-WZ series three channel universal input paperless recorder

parameter
Category
Specification
input signal
Input channel
Route 1-3
analog quantity
Thermocouples: K, B, S, T, E, J;
Thermistor: Pt100, Cu50
DC current: 0-10mA, 4-20mA

DC voltage: 0-20mV, 0-10mV, 0-5V, 1-5V;

output signal
Analog output
DC 4~20mA
alarm output
2-channel configurable relay contact output, contact capacity AC250V/3A (resistive load) default to normally open contact

Communication output

Communication interface: RS232C or RS485 baud rate -1200bps 2400bps 960ps 19200bps
Printing interface: RS232C serial micro printer
Feed output
DC24V, Load current ≤ 65mA
Other functions
measurement accuracy
± 0.5% FS or ± 0.2% FS (special customization required)
resolution
± 1 word
sampling period
1 second
Recording interval
1. Nine levels of 2, 4, 6, 15, 30, 60, 120, and 240 seconds are available for selection
Record length
3 days/pen (recording interval of 1 second) to 720 days/pen (recording interval of 4 minutes)
Recording interval
1 second
2 seconds
5 seconds
10 seconds
15 seconds
30 seconds
1 point
2 points
4 points
Record length
3 days

6 days

15 days

30 days
45 days
90 days
180 days
360 days
720 days
display range
-9999~19999
display mode
128 × 64 dot matrix LCD display, LED backlight, display content can be composed of Chinese characters, Western languages, numbers, process curves, light columns, etc. Page turning can be completed through panel buttons, historical data can be searched before and after, and curve time scale can be changed
Parameter Settings
Chinese menu prompt, panel button settings or communication settings on the upper computer
protection method
FLASH stores settings parameters and historical data, and the data is permanently saved after power failure. Using hardware clock, it can run accurately even after power failure
Usage environment
Ambient temperature
0~50℃
relative humidity
≤85%RH, Avoid strongly corrosive gases
power supply voltage
AC85-265V (Switching Power Supply)
DC24 ± 2V (switching power supply)
power waste
<10W

Selection code

PC-
Puchen Company's business logo

serial number

WZ-C81 Single circuit 160 × 80mm (horizontal)

WZ-C82

Dual circuit 160 × 80mm (horizontal)

WZ-C83

Three circuit 160 × 80mm (horizontal)

Alarm selection

one

Without alarm

three

With alarm

contact method

0

No communication

-2

RS-232C

-3

RS-232C printing port

-8

RS-485

Transmission output mode

0

No transmission output

one

4-20mA transmission output

Alarm point selection

0

No alarm control output

-1

A little alarm output

-2

Two point alarm output

Feed output

-P

One DC24V output

power supply method

-T AC85-265V (Switching Power Supply)
-W DC24V (direct current)

Precision selection

-
Not selected, with an accuracy of 0.5 level
B Accuracy level 0.2

Additional feature selection

/U

Comes with USB transfer function


Note:
When the input channel is two-way, it is allowed to have two transmission outputs. If the input signal is three-way, only one output signal can be carried.
The first and second signals can be switched, while the third signal can only be 4-20mA and 1-5V.

Instrument appearance and opening size

PC-WZ Series Three Channel Universal Input Paperless Recorder
Product Category
All product information on this website is translated by Baidu machine translation, which may not be completely accurate and is only for reference