DCSG-2099 Multi parameter AnalyzerDetailed product description:Column tabs
DCSG-2099 multi parameter industrial water quality analyzer, capable of simultaneously measuring: conductivity TDS、 Resistivity, temperature pH、ORP、 Dissolved oxygen, dissolved oxygen saturation, residual chlorine, a total of nine parameters. Each channel is independent of each other, without switching switches, and does not interfere with each other.

Key Features:
Chinese menu style operation: adopting a menu structure, similar to microcomputer operation, easy to operate, with Chinese prompts throughout the operation steps, and can be easily completed without the need for an instruction manual.
Multi parameter display on the same screen: Display conductivity, temperature, pH value on the same screen ORP、 Dissolved oxygen, hypochlorous acid, or residual chlorine. It can also switch to display the 4-20mA current value and electrode signal corresponding to each parameter.
Current isolation output: Six independent 4-20mA currents, using photoelectric coupling isolation technology, strong anti-interference ability, and can be transmitted remotely.
RS485 communication interface: It can be easily connected to a computer for monitoring and communication.
Manual current source function: It can check and set the output current value arbitrarily, making it convenient to detect the recorder and lower computer.
Automatic temperature compensation: Automatic temperature compensation within 0-99.9 ℃.
Waterproof and dustproof design: protection level IP65, suitable for outdoor use.

Technical specifications of DCSG-2099 multi parameter analyzer
1. Display mode: Chinese LCD display, Chinese menu style operation.
2. Conductivity: electrode constant × (1-3000) μ S/cm, 0-99.9 ℃.
0.01~30μS/cm; (0.01 electrode); 0.1~300μS/cm; (0.1 electrode);
1.0~3000μS/cm; (1.0 electrode); 10~30000μS/cm; (10.0 electrode).
3. Conductivity accuracy: ± 0.5% FS, ± 0.3 ℃.
4. PH: Measurement range: (0.00~14.00) pH; Basic error of electronic unit: ± 0.02pH.
5. The basic error of the instrument is ± 0.05pH.
6. Temperature: Measurement range: (0.0~100.0) ℃; Basic error of electronic unit: ± 0.3 ℃.
7. Basic instrument error: 0.5 ℃ (0.0 ℃≤ T ≤ 60.0 ℃); Other ranges up to 1.0 ℃.
8. ORP: Measurement range: (-1999~1999) mV; Basic error of electronic unit: ± 1mV.
9. The basic error of the instrument is ± 5mV.
10. Dissolved oxygen: Measurement range: 0-100.0 ug/L, 0.00-20.00mg/L, 0.0-200.0%.
11. Basic error of the whole machine: ug/L: ± 1.0% FS; mg/L: ± 0.5% FS, temperature: ± 0.5 ℃.
12. Residual chlorine: Measurement range: 0-20.00mg/L; Resolution of 0.01mg/L.
13. Residual chlorine accuracy: ± 2% or ± 0.035mg/L, whichever is greater.
14. TDS: Measurement range: 0.000-999.9mg/L, 1.000g/L-19.99g/L.
15. Basic error of electronic unit: ± 1.0% (FS); Basic instrument error: ± 2.0% (FS).
16. Electrical resistivity: Measurement range: 0-20.00 M Ω• cm.
17. Basic error of electronic unit: ± 1.0% (FS); Basic instrument error: ± 2.0% (FS).
18. Each channel is independent of each other, and the data of each channel is measured simultaneously.
19. Conductivity, temperature pH、ORP、 Dissolved oxygen and residual chlorine are displayed on the same screen, while other displays can be switched.
20. Current isolation output: Each parameter is independently 4-20 mA (load<750 Ω).
21. Power supply: AC220V ± 22V, 50Hz ± 1Hz, optional DC24V ± 2.4V.
22. RS485 communication interface (optional).
23. Protection level: IP65.
24. Working conditions: ambient temperature of 0-60 ℃, relative humidity ≤ 90%.
25. Weight and dimensions of the entire instrument set: 40.0 kg, 144 mm (width) x 144 mm (height) x 110 mm (thickness).