UQZ-58D series top mounted magnetic float level gaugeDetailed product description:Column tabs

Overview
The UQZ-58D series top mounted magnetic float level gauge (hereinafter referred to as the level gauge) is used to measure the liquid level of the medium in the container during the production process. It is a top mounted multifunctional liquid level gauge that can be used for on-site indication, remote digital display, alarm and control, as well as for corrosive media and explosion-proof working environments. Equipped with a remote transmitter, it outputs 4-20mADC standard electrical signals and can be matched with DDZ-III electric unit combination instruments for industrial production automation control systems. It is widely used in industrial sectors such as petroleum, chemical, pharmaceutical, and light industry.
The liquid level gauge is manufactured according to the HG/T21584-95 standard of the Ministry of Chemical Industry (UHZ series indicating magnetic float liquid level gauge).

characteristic
The structure is simple and easy to install and use.
Large measurement range and multiple functions.
Corrosion resistant and explosion-proof.

Main technical parameters of UQZ-58D series top mounted magnetic float level gauge
Basic on-site indicating liquid level gauge
Measurement range: 0-6000mm.
Indication error: ± 10mm.
Rated working pressure: 0.6, 1.6, and 2.5 MPa.
Nominal diameter: DN150mm.
Medium temperature for use: -40 to+250 ℃.
Medium density: not less than 0.45g/cm3.
Remote transmission device
Measurement range: 0-6000mm.
Output signal: 4-20mADC.
Basic error: ± 10mm.
Power supply: 24VDC.
Load power supply: 250 Ω (excluding wire resistance of 200 Ω).
Maximum transmission distance: 3Km.
Explosion proof type and marking: 1. IB Ⅱ CT4 intrinsic safety type; 2. D Ⅱ BT4 explosion-proof type.
Associated device: LB830 Zener Safety Manual (for intrinsically safe transmitters).
alarm
1. Power supply voltage: 1.24VDC 2.20VAC, 50Hz
2. Output signal: Two normally open, two normally closed switch signals
3. Switch contact capacity: 220V, 2A
4. Working life:>105 times
5. Alarm accuracy: ± 5mm
6. Environmental temperature: -40 to+60 ℃
7. Environmental vibration: frequency ≤ 25Hz, amplitude ≤ 0.5mm
8. Explosion proof type and explosion-proof mark: Passed the inspection of the Electrical Instrument Explosion proof Quality Supervision Center of the Ministry of Chemical Industry. Explosion proof mark dIIBT4.
9. Explosion proof hose interface for alarm switch: M20 × 1.5 internal thread.
structural principle
The basic liquid level gauge cylinder is equipped with a magnetic float, which moves up and down with the rise and fall of the liquid in the container, and uses magnetic coupling transmission to synchronously drive the liquid level indicator, thereby achieving on-site indication of the liquid level.
The total resistance value of the measurement circuit of the remote transmitter changes correspondingly with the magnetic induction effect of the magnetic float that rises and falls with the liquid level, and after amplification and processing by the electronic module, it is converted into a 4-20mADC standard electrical signal and transmitted remotely to the secondary instrument (indicating, recording or controlling instrument).
After block amplification and other processing, it is converted into a 4-20mADC standard electrical signal and transmitted remotely to secondary instruments (indicating, recording or controlling instruments).
The alarm is driven by the magnetic coupling transmission effect, which drives the follower components in the alarm to activate the proximity switch to achieve liquid level alarm and control.
Overall dimensions
The basic liquid level gauge is equipped with a liquid level remote transmitter and alarm, as shown in Figure 1 for its appearance and dimensions.
Install
The basic level gauge adopts JB/T82.2-94, DN150mm, PN1.6MPa or PN2.5MPa flat flange installation. The flat flange with the same standard and specification is configured by the user on the top of the process vessel and storage tank.
Wire entry connector for remote transmitter and alarm: M20 × 1.5 internal thread connector.
Ordering Instructions
Please provide detailed information on the model, name, measurement range, medium name and density, rated working pressure and temperature, and number of ordered liquid level gauges.
If users have special requirements (such as corrosion-resistant materials, etc.), they should be explained.
1. Remote transmitter
2. Indicator panel level gauge
3. Upper limit alarm
4. Lower limit alarm
5. Flanges

UQZ-58D series top mounted magnetic float level gauge with external dimensions

UQZ-58D series top mounted magnetic float level gauge
Product Category